EasyManua.ls Logo

Siemens MOBILETT XP Digital - Page 32

Siemens MOBILETT XP Digital
74 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
32 General maintenance information
MOBILETT XP Digital SPR8-230.831.30.03.02 Siemens AG
11.05 CS SD 24
Page 32 of 74
Medical Solutions
2.3 Are undamaged accessories being
used?
3.1.1 Damage, check:
SIM DAP (optional)
SIM Remote control (optional)
2.4 System radiation protection devices
present and undamaged? (No on-site
radiation shield)
3.3.3 Check of the radiation indicator,
checks:
SIE Radiation indicator
SIE Ready indicator
SIE Acoustic signal
3.3.4 Manual termination of exposure,
checks:
SIE "ERR 39" display
SIE Acoustic signals
2.5 Are the required operator documents
complete, present and legible?
3.1 Visual inspection, check:
SI Customer documentation
Ask the customer for additional required
operator documents and check them for
completeness, availability, and legibility.
2.6 Are all warning labels in place and
recognizable?
3.1 Visual inspection, check:
SIM Signs
2.7 Are the operating symbols, light indi-
cators and button labels OK?
3.3.2 Control panel displays, checks:
SIE kV/mAs 7-segment displays
SIE kV/mAs displays according to operating
instructions
3. Electrical checks n.a.
3.1 Measurement of ground wire resis-
tance
3.10 Protective conductor test, check:
SIE Protective conductor test
3.2 Leakage current or equivalent leak-
age current measurement
n.a.
3.2.1 Leakage current measurement 3.11 Leakage current measurement, check:
SIE Leakage current measurement
3.2.2 Equivalent leakage current mea-
surement
n.a.
3.3 Patient leakage current or equivalent
patient leakage current measurement
n.a.
Test report/
Technical safety checks
Maintenance instructions for the
check/comments