36 Inspection and maintenance
MOBILETT XP Digital SPR8-230.831.30.03.02 Siemens AG
11.05 CS SD 24
Page 36 of 74
Medical Solutions
3-
3 Inspection and maintenance
3.1 Visual check
SIM Signs
Check all signs on the system with the model and serial numbers as well as warnings.
The placement of these signs is described in the operating instructions ("Signs" chap-
ter).
Verify that all signs are present and legible.
SI Customer documentation
The customer must be able to provide the following documents. The documents are to
be checked for availability, completeness, and legibility.
- Operating instructions
- Start-up protocol
- Last maintenance protocol
- Quality assurance protocol
3.1.1 Damage
Check the unit for external damage (cracks, breaks, scratches, corrosion etc.).
Check whether the single tank or the collimator show any signs of mechanical defects
that can impair radiation protection.
Repair or replace defective parts.
SIM Covers
SIM Detector and detector holder
Check the condition of the detector side covers and replace them if necessary.
SIM Control panel and display panel
SIM Release cable
SIM Single tank
SIM Collimator
SIM DAP (optional)
The DAP ionization chamber should not have any scratches.
No damage to DAP supply line, plug or display
SIM Remote control (optional)
No damage to remote control
- Maintenance records
- Service reports for repairs and adjustments
- DHHS documentation