Siemens AG SPR8-230.831.30.03.02 MOBILETT XP Digital
11.05 CS SD 24
Inspection and maintenance 47
Page 47 of 74
Medical Solutions
3.2.9 Power cable
NOTE
Replacement of cable winch including cable:
MOBILETT XP Digital every 4 years
PMP Cable winch replacement
If the replacement interval has been reached, replace the cable winch according to the
"Replacement of parts" instructions.
SIM Checking the power cable
Pull the entire power cable (6m) out of the cable winch. Check the cable for damage.
SIM General cable winch information
Pull the entire power cable (6 m) out of the cable winch. Hold the cable while coiling and
move it up and down until the cable is fully coiled. Verify that the cable winch and the
brake work properly.
PMP Cleaning
Clean the power cable with a cloth. Dampen the cloth with a mixture of lukewarm water
and household cleaner.
3.2.9.1 Power plug
SIM Checking the power plug
Check the power plug for proper functioning.
Open the plug and check the terminals and strain relief.
¹ The ground wire has to be slightly longer than the other wires (applies to all plug
types).
Fig. 14: Ground wire
3.2.10 Lubrication
PMP Pull-bar slide bushing