Siemens AG SPR8-230.831.30.03.02 MOBILETT XP Digital
11.05 CS SD 24
Inspection and maintenance 49
Page 49 of 74
Medical Solutions
3.3 Function inspection
3.3.1 Operating data
Activating service programs
• Switch off the system, unplug the power plug, remove the top cover and place it side-
ways on a table or the like. Do not disconnect any cables.
• Move service switch SW2B into position 2 (service) board D916 (CPU) (Fig. 16 / p. 49).
• - Switch the system on.
¹ "Pr 1" is displayed on the control panel.
• Use the mAs+/mAs- buttons to select the desired service program.
NOTE
The X-ray function is deactivated during an active service pro-
gram.
For detailed information, see Troubleshooting Instructions
SPR8-230.840...
Fig. 16: CPU D916
3.3.1.1 Reading out the operating information
PMF Exposure counter
• Select ServiceProgram 02 and record the "exposure" counter status in the protocol.
PMF Error memory
• Select ServiceProgram 03 and read out the error memory. If a defective component
message is displayed, check it according to the troubleshooting instructions. Reset the
memory in service program 04.
• Switch off the system.
• Return service switch SW2B D916 (CPU) to position 1 (user).