Siemens AG SPR8-230.831.30.03.02 MOBILETT XP Digital
11.05 CS SD 24
Inspection and maintenance 49
Page 49 of 74
Medical Solutions
3.3 Function inspection
3.3.1 Operating data
Activating service programs
Switch off the system, unplug the power plug, remove the top cover and place it side-
ways on a table or the like. Do not disconnect any cables.
Move service switch SW2B into position 2 (service) board D916 (CPU) (Fig. 16 / p. 49).
- Switch the system on.
¹ "Pr 1" is displayed on the control panel.
Use the mAs+/mAs- buttons to select the desired service program.
NOTE
The X-ray function is deactivated during an active service pro-
gram.
For detailed information, see Troubleshooting Instructions
SPR8-230.840...
Fig. 16: CPU D916
3.3.1.1 Reading out the operating information
PMF Exposure counter
Select ServiceProgram 02 and record the "exposure" counter status in the protocol.
PMF Error memory
Select ServiceProgram 03 and read out the error memory. If a defective component
message is displayed, check it according to the troubleshooting instructions. Reset the
memory in service program 04.
Switch off the system.
Return service switch SW2B D916 (CPU) to position 1 (user).