50 Inspection and maintenance
MOBILETT XP Digital SPR8-230.831.30.03.02 Siemens AG
11.05 CS SD 24
Page 50 of 74
Medical Solutions
3.3.2 Displaying the control panel
PMF Mains display
• The yellow line voltage control LED must be lit when the line voltage plug is plugged in
(3/Fig. 17 / p. 50).
• System ON.
PMF kV/mAs default values
SIE kV/mAs segment displays
• The kV/mAs default values must display and be completely legible (take the default val-
ues from the start-up protocol).
PMF Ready indicator
• The green ready LED (2/Fig. 17 / p. 50) should light up within 15 seconds of "system
ON".
SIE kV/mAs displays according to operating instructions
• Select all kV/mAs values.
3.3.3 Checking the radiation indicator
Fig. 17: Control and display
SIE Radiation indicator
SIE Ready indicator
SIE Acoustic signal
• System is switched on.
• Set 81 kV and 10 mAs.
X
• Initiate an exposure.
The radiation indicator (1/Fig. 17 / p. 50) lights up during the exposure; at the same
time an acoustic signal sounds and the "ready" light goes out.