Siemens AG SPR8-230.831.30.03.02 MOBILETT XP Digital
11.05 CS SD 24
Inspection and maintenance 51
Page 51 of 74
Medical Solutions
• The "ready" light (2/Fig. 17 / p. 50)goes on again within a maximum of 15 seconds.
3.3.4 Manual termination of exposure
The user must be able to cancel an exposure at any time.
SIE “ERR 39” display
SIE Acoustic signals
• System is switched on.
• Set a midrange kV value (e.g., 70 kV) and the highest possible mAs value.
X
• Release an exposure and interrupt it immediately.
¹ The display then reads "ERR 39" and several short acoustic signals are emitted
("ERR 39 = exposure aborted). Acknowledge the error message with the "colli-
mator light" key on the control panel.