58 Inspection and maintenance
MOBILETT XP Digital SPR8-230.831.30.03.02 Siemens AG
11.05 CS SD 24
Page 58 of 74
Medical Solutions
3.5 Battery and motor drive inspection
3.5.1 Batteries
NOTE
The life of battery packs BK1-BK4 is estimated at 12-24 months.
There is no acceptable measuring procedure to determine the bat-
tery quality.
Battery life is dependent on the charging and discharging times
which are largely determined by the user.
The user must provide the charge/discharge time ratio.
The longest charging time is approx. 12 hrs. The three LEDs light up yellow on the control
panel (3/Fig. 27 / p. 58).
One discharge cycle allows for 80 exposures with maximum kV/mAs values and transport
operation of approx. 3,000 m.
Realistic statement: The MOBILETT does not need to be charged during the 12-hour
workday.
If the minimum battery charge is reached in < 5 hrs., the battery must be replaced.
Fig. 27: Battery status display
PMF Charging performance
Ask operating personnel about the charging cycles.
¹ If necessary, replace the batteries as described in the "Replacement of parts"
instructions.
3.5.2 Motor drive
NOTE
Completely reassemble the system and attach all the covers for
the following work procedures.
PMF Forward/backward
PMF Slow/fast
Turn the system ON and release the hand brake.
Check the transport movements according to the operating instructions.