Inspection and maintenance 61
Page 61 of 74
Medical Solutions
Fig. 28: kV measurement
X • Set the following exposure parameters and initiate an exposure after each setting:
PMF kV accuracy 52 kV, 50 mAs
PMF kV accuracy 81 kV, 20 mAs
PMF kV accuracy 133 kV, 12.5 mAs
Record the measured values.
Oscilloscope method:
"KVS" measuring point for the nominal value and “KV” measuring point for the actual
value. The measuring ratio is 30 kV/V.
• Open the top system cover and place the cover sideways on a table or the like. Do not
disconnect any cables.
kV meter
Select
Exposure parameters
Limit values
DC voltage
52 kV 50 mAs 49.0 - 55.0 kV
81 kV 20 mAs 77.0 - 85.0 kV
133 kV 12.5 mAs 126.4 - 139.6 kV