42 Error messages
MOBILETT XP SPR8-230.840.01.05.02 Siemens AG
11.05 CS SD 24
Page 42 of 72
Medical Solutions
ERR 31 kV exceeds permissible limit
System
mode:
Exposure
Explanation: D916 (CPU) detects max. kV.
Action:
Check the test point "overvoltage error" TP - KVM on D916 (CPU), (high
= error) after exposure.
Check the cable connection between the single tank and D916.
Check the highest frequency of the main inverter, test point FC2 on
D916 (CPU).
Check the filament current.
Possible defects: D916, D907; single tank.
ERR 34 Tube current exceeds permissible limit
System
mode:
Exposure
Explanation: D916 (CPU) detects I
JR
max. during an exposure (test point
D916, TP JIM).
Action:
Check test point TP - JIM on D916 (CPU), (high = error).
Check the filament current with preparation. Set potentiometer P5 (fila-
ment current during preparation).
¹ Sporadic error message - check the highest frequency of the fila-
ment inverter.
Possible defects: D916, D907.
ERR 35 Short-circuit at the main inverter
System
mode:
Exposure
Explanation: D916 (CPU) detects a main inverter short-circuit.
Action:
Check test point TP - SCM on D916 (CPU) after exposure, (high = er-
ror).
ERR 35 with exposures in all kV ranges indicates a probable single tank
defect.
Check the cable connections between D916 and D962 (X20).
Check the cable connections between D962 and the single tank (U - V).
Check C1.