42 Error messages
MOBILETT XP SPR8-230.840.01.05.02 Siemens AG
11.05 CS SD 24
Page 42 of 72
Medical Solutions
ERR 31 kV exceeds permissible limit
System
mode:
Exposure
Explanation: D916 (CPU) detects max. kV.
Action:
• Check the test point "overvoltage error" TP - KVM on D916 (CPU), (high
= error) after exposure.
• Check the cable connection between the single tank and D916.
• Check the highest frequency of the main inverter, test point FC2 on
D916 (CPU).
• Check the filament current.
Possible defects: D916, D907; single tank.
ERR 34 Tube current exceeds permissible limit
System
mode:
Exposure
Explanation: D916 (CPU) detects I
JR
max. during an exposure (test point
D916, TP JIM).
Action:
• Check test point TP - JIM on D916 (CPU), (high = error).
• Check the filament current with preparation. Set potentiometer P5 (fila-
ment current during preparation).
¹ Sporadic error message - check the highest frequency of the fila-
ment inverter.
Possible defects: D916, D907.
ERR 35 Short-circuit at the main inverter
System
mode:
Exposure
Explanation: D916 (CPU) detects a main inverter short-circuit.
Action:
• Check test point TP - SCM on D916 (CPU) after exposure, (high = er-
ror).
• ERR 35 with exposures in all kV ranges indicates a probable single tank
defect.
• Check the cable connections between D916 and D962 (X20).
• Check the cable connections between D962 and the single tank (U - V).
• Check C1.