7 Operation at Use of a PC
7.5 Maintenance
193SICAM P850/P855, 7KG85xx, Device Manual
E50417-H1040-C482-A6, Edition 10.2018
Click Browse....
The Choose file dialog opens.
Select the desired update (extension .pck) in the directory.
Click Open.
The selected path is inserted in the input/output window, Figure 7-70, into the Browse... field.
Click Open on the tab.
After approx. 2 s, the message Action was successful! is displayed in the input/output window.
Device firmware, default set of parameters, text libraries, HTML files or parts thereof are uploaded within
one minute.
The device then restarts automatically.
NOTE
If you want to carry out a firmware update, you must stop the record manually before., see Recording
Parameters and Power Quality Data.
Do not switch off the supply voltage during the upload process as this can lead to data loss.
NOTE
For SICAM P850 firmware uploading to V2.20, refer to PI_FW-Update_SICAM_P85x_V2.00-to-V2.20,
en_US.pfd.