Adjustments / Programs 5 - 5
Siemens AG Register 5 RXR2-130.061.02 Page 5 of 22 SIREMOBIL Compact
Medical Engineering Rev. 03 10.03 CS PS 24 System Manual
Dose rate adjustment 5
Prerequisite 5
The "Adjustment mode" parameter in the "Dose rate adjustment" parameter group must
be programmed to "Semi automatic".
Preparations 5
Switch the SIREMOBIL Compact on.
Connect the service PC to the serial interface of the SIREMOBIL Compact.
Start the service program.
Select the “Diagnostics...”menu.
Select the “Dose rate adjustment” menu in the Combo Box “Group 1”.
- The subsequent display is visible on the service PC:
Insert 25mm AL and 2.1mm Cu filters at the radiation exit port of the Sirephos.
The grid remains on the I.I.
X-ray radiation!
See Chapter 1, Safety Information.
Protect yourself against radiation exposure. Wear a lead apron.
- AUTO REGULATION
indicates the current status of the automatic adjustment
(active / not active / done)
- BRIGHTNESS DEVIATION
indicates the difference of the brightness actual value minus the
brightness nominal value
- CONTROL CURVE
indicates the currently selected fluoroscopic curve
-ADJUSTMENT
Indicates the operating control mode
(Generator / Camera Iris / Stop)
- BV-ZOOM STATUS
indicates the I.I. format (survey / zoom).
-DOSERATE
indicates the selected dose control level (LOW / MID / HIGH)
Only the MID dose control level is used
Correction factors:
Grid factor 23cm (9”) I.I. and 17cm (7”) I.I.: 1.5
Chamber correction factor for Dali and Nomex: 1.06
Dose rate to be set at the I.I. input screen:
23 (9”) I.I. and 17cm (7”) I.I.: 220 nGy/s
WARNING