Adjustments / Programs 5 - 11
Siemens AG Register 5 RXR2-130.061.02 Page 11 of 22 SIREMOBIL Compact
Medical Engineering Rev. 03 10.03 CS PS 24 System Manual
• Release fluoroscopy. The X-Iris must close between 4 - 4.5 cm.
• Stop fluoroscopy.
• Select "5. CLOSED POSITION" on the service PC and click on "Execute".
• Close the status window by clicking on the OK button.
• For the query "Do you want to reset the unit," click "No".
• Select "8. Init Off & Store Positions" on the service PC and click on "Execute".
• Close the status window by clicking on the OK button.
• For the query "Do you want to reset the unit," click "Yes".
• Switch the SIREMOBIL Compact OFF and ON again.
• Check the X-Iris again in survey format and zoom format. The collimator blades must still
be visible on the edge of the image. If necessary, repeat the adjustment.
• Stop fluoroscopy.
• Return to the main menu on the service PC with the ESC key.
• Select the Data, Backup... menu.
• Select Parameter in the backup window.
• Click on "Backup".
• At "Remarks for Parameter" enter "X-Iris set“.
• Click on "OK". The new position values for the collimator will be stored on the disk.
• Return to the main menu with the ESC key.