SIREMOBIL Compact Register 5 RXR2-130.061.02 Page 14 of 22 Siemens AG
System Manual Rev. 03 10.03 CS PS 24 Medical Engineering
5 - 14 Adjustments / Programs
Click "OK". The new position values for camera rotation will be stored to the disk.
Switch the SIREMOBIL Compact OFF and ON again.
Check the adjustments again. In the 0° position, the object image must correspond to the
one in Fig. 4.
Calibrating the area dose product measurement device 5
Calculating the CONSTANT parameter (Qt * Kg) 5
Read and record the Kg value shown on the electronics unit of the area dose product
measurement device.
Read the Qt value on the dose chamber of the area dose product measurement device.
Multiply these values and record them. The resulting value is the CONSTANT (QT*Kg)
value to be programmed.
Programming 5
Connect the service PC to the serial interface on board D1.
Start the service program.
Select the Adjustment, Parameter menu.
In the Combo box "Parameter Groups", select "Diamentor" .
Click on "Get from Unit".
The Parameter "1. Diamentor Existing" must be on YES .
Select parameter "2. Constant (Qt * Kg)" .
In "Value Actual:" field, enter the parameter you have calculated.
Click on "Set Value".
Click on "Put to Unit" .
Click on "Save to File".
This adjustment is used to calibrate the area dose product mea-
surement device of the SIREMOBIL Compact display.
You must exit the service program with "Logoff" and "Quit" to
store the modified parameters to disk as a new file.
NOTE
NOTE