Adjustments / Programs 5 - 15
Siemens AG Register 5 RXR2-130.061.02 Page 15 of 22 SIREMOBIL Compact
Medical Engineering Rev. 03 10.03 CS PS 24 System Manual
Checking the programmed values 5
Prerequisites
The Qt * Kg value was checked and is programmed. Refer to Programming.
Connect the service PC to the serial interface on board D1.
Start the service program.
Select the Adjustment, Calibrations menu.
Click "Yes" to answer the query.
In the Combo box "Function Group:", select "Diamentor".
In the window "Available Functions", select the line "Start Diamentor Test" and click on
"Execute" .
The tested Qt*Kg value is displayed on the monitor of the SIREMOBIL Compact.
This value must match the programmed CONSTANT (Qt*Kg) value.
Tolerance: +/- 5%.
By pressing the -0- key (reset key for fluoroscopic time) the test is repeated and is
displayed again on the monitor.
The potentiometer CAL on the electronic system of the area dose product measurement
device can be used to adjust the test value to the calculated and programmed Qt*Kg
value.
After the adjustment, select "Stop Diamentor Test" on the service PC and click on
"Execute".
The test is ended.
Check the accuracy of the area dose product measurement device. Refer to the next
page.
Checking the programmed values will activate the self-test in the
electronic system of the area dose product measurement device.
This serves to recalibrate the electronic system of the area dose
measurement device.