Adjustments / Programs 5 - 15
Siemens AG Register 5 RXR2-130.061.02 Page 15 of 22 SIREMOBIL Compact
Medical Engineering Rev. 03 10.03 CS PS 24 System Manual
Checking the programmed values 5
Prerequisites
The Qt * Kg value was checked and is programmed. Refer to Programming.
• Connect the service PC to the serial interface on board D1.
• Start the service program.
• Select the Adjustment, Calibrations menu.
• Click "Yes" to answer the query.
• In the Combo box "Function Group:", select "Diamentor".
• In the window "Available Functions", select the line "Start Diamentor Test" and click on
"Execute" .
• The tested Qt*Kg value is displayed on the monitor of the SIREMOBIL Compact.
• This value must match the programmed CONSTANT (Qt*Kg) value.
Tolerance: +/- 5%.
• By pressing the -0- key (reset key for fluoroscopic time) the test is repeated and is
displayed again on the monitor.
• The potentiometer CAL on the electronic system of the area dose product measurement
device can be used to adjust the test value to the calculated and programmed Qt*Kg
value.
• After the adjustment, select "Stop Diamentor Test" on the service PC and click on
"Execute".
• The test is ended.
• Check the accuracy of the area dose product measurement device. Refer to the next
page.
Checking the programmed values will activate the self-test in the
electronic system of the area dose product measurement device.
This serves to recalibrate the electronic system of the area dose
measurement device.
NOTE