SIREMOBIL Compact Register 5 RXR2-130.061.02 Page 18 of 22 Siemens AG
System Manual Rev. 03 10.03 CS PS 24 Medical Engineering
5 - 18 Adjustments / Programs
Evaluation 5
Calculate the deviation between the measured area dose product and the displayed area
dose product:
The deviation must be < 0.3.
Deviation = [(displayed area dose product FDPa) - (measured area dose product FDPg)] /
(measured area dose product FDPg)
For deviations > 0.3, the area dose product measurement device must be replaced.