Replacing boards / replacing components 6 - 9
Siemens AG Register 5 RXR2-130.061.02 Page 9 of 18 SIREMOBIL Compact
Medical Engineering Rev. 03 10.03 CS PS 24 System Manual
Checks and adjustments 6
• Check the I.I. electrode voltages according to the test protocol for the I.I.
-Do not readjust in cases where there are only slight deviations from the values on the
test protocol (measurement device tolerances).
• Check that the camera optics are centered to the I.I. output and adjust them if necessary;
refer to "Replacing the camera optics ", subitem "Checking the centering of the camera
optics to the I.I. output " and "Centering the camera optics to the I.I. output".
• Check the camera rotation and readjust it if necessary.
• Check the setting of the radiation iris and readjust it if necessary.
• Check the dose rate and readjust it if necessary.
• Perform the IQ Quick test.
- When doing this, check the resolution first and, if necessary, readjust the optical
resolution of the Compact optics.
• Complete the country-specific acceptance. (§16 Partial acceptance... /DHHS...)
Replacing the compact optics 6
Removing the Compact optics 6
• Turn the C-arm so that the I.I. is at the top.
• Loosen and remove the I.I. cover.
• Rotate the camera in the "0" position (6/Fig. 6).
Electrical voltage!
See Chapter 1, Safety Information.
Prior to removing the mini voltage supply, the system must be
switched off for at least 3 minutes until the high voltage in the
system and in the I.I. mini voltage supply dissipates.