SIREMOBIL Compact Register 5 RXR2-130.061.02 Page 16 of 18 Siemens AG
System Manual Rev. 03 10.03 CS PS 24 Medical Engineering
6 - 16 Replacing boards / replacing components
• Configure the following parameters in User setup for Memoskop:
• Perform the IQ quick test.
A Operating mode parametersChange the programmed parameters only upon the cus-
tomer’s request.
B Video recorder output Configure to monitor A.
C Documentation For SIREMOBIL Compact with 1 monitor:
configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
Configure to monitor B.
D Text monitor For SIREMOBIL Compact with 1 monitor:
configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
configure to both monitors.
E Monitor split Configure horizontally. Configure vertically only upon cus-
tomer request.
F Time Configure Yes.
G Date / Time set Set the current date and time.
H Hospital name Program the hospital.
I Dose storage If area dose product meter is present, (Diamentor) config-
ure Yes. Otherwise configure No.
K Dose documentation If area dose product meter is present, (Diamentor) config-
ure yes. Otherwise configure No.