Loading...
Installation of TFT displays 2 - 7
Siemens AG SPR2-230.812.02 Page 7 of 8 SIREMOBIL
Medical Solutions Rev. 02 02.06 CS PS SP System Manual
Wiring of the displays 2
Open the rear panel of both TFT displays above the connections.
Route the power line cables and the video cables to the TFT displays.
BARCO displays 2
Connect the video cable to the BNC sockets VIDEO of the displays and lock it.
Connect the shieldings of the video cables to the ground of the display.
Plug in the power cables.
Close both covers again located above the display connections.
Running the cables 2
If required, retract the power line cables and video cables into the monitor trolley.
- Route the cables without sharp bending radius.
- It is recommended to ensure that the cables are covered by both support arm rails in
the area between support arm base and the displays (see Fig. 11).
- The cables should rest in the cable bushes of the support arm base without excessive
pulling force or stress (Fig. 15).
Reattach the cover of the support arm (Fig. 15).
When present, switch on the power switches of the TFT displays.
Fig. 15 Cover of the support arm
Mark for the keyboard tray