ID Symbols Message Cause/Remedy
A7 Event counter 2
Number overruns above
threshold
Maintenance alarm
The number of overruns of the process value (set in parameters "Upper limit"
and "Monitored value") has reached the threshold.
Reset and acknowledge event counter.
Check process conditions.
Check limit monitoring and event counter settings.
A8 Event counter 2
Number underruns above
threshold
Process value alarm
The number of underruns of the process value (set in parameters "Lower
limit" and "Monitored value") has reached the threshold.
Reset and acknowledge event counter.
Check process conditions.
Check limit monitoring and event counter settings.
A9 Event counter 2
Number underruns above
threshold
Maintenance required
The number of underruns of the process value (set in parameters "Lower
limit" and "Monitored value") has reached the threshold.
Reset and acknowledge event counter.
Check process conditions.
Check limit monitoring and event counter settings.
AA Device lifetime: Mainte‐
nance demanded
Forthcoming end of congured device's lifetime.
Maintenance is strongly recommended as soon as possible.
Ab Device lifetime: Mainte‐
nance required
Forthcoming end of congured device's lifetime.
Maintenance of device should be planned.
AC Sensor lifetime: Mainte‐
nance demanded
Forthcoming end of congured sensor's lifetime.
Maintenance is strongly recommended as soon as possible.
Ad Sensor lifetime: Mainte‐
nance required
Forthcoming end of congured sensor's lifetime.
Maintenance of device should be planned.
AE Service: Maintenance de‐
manded
Forthcoming end of the congured service interval.
Maintenance is strongly recommended as soon as possible.
AF Service: Maintenance re‐
quired
Forthcoming end of the congured service interval.
Maintenance of device should be planned.
Diagnostics and troubleshooting
4.1Diagnostic messages
SITRANS P320/P420
14 Service Manual, 11/2023, A5E44914141-AC