5.1.6 Commissioning the device after transmitter electronics have been replaced
Procedure
1. Commission the device according to the operating instructions.
2. In order for the measuring cell data to be read by the new measuring cell electronics, restore
the factory setting on the device.
Use the "RESET" parameter [35] > "FACT" via local operation.
3. If required, write the saved user data to the device via remote operation (such as SIMATIC
PDM).
Result
The device is now ready for operation.
5.2 Replacing the measuring cell
Procedure
Note
When replacing the measuring cell, observe the relevant certicates (e.g. oxygen applications,
material certicates).
To replace the measuring cell, carry out the following steps:
1. If required, back up the user data via remote operation (e.g. SIMATIC PDM).
2. Remove the display.
Removing the display (Page23)
3. Remove the transmitter electronics.
Removing the transmitter electronics (Page24)
4. Remove the current measuring cell.
Removing the measuring cell (Page28)
5. Install the new measuring cell.
Installing the measuring cell (Page30)
6. Install the display.
Installing the display (Page26)
7. Commission the device.
Measuring accuracy following measuring cell replacement
After you have replaced the measuring cell, the measuring accuracy of the device may be
outside the specication.
To comply with the specied measurement accuracy, carry out a sensor calibration.
Replacing spare parts
5.2Replacing the measuring cell
SITRANS P320/P420
Service Manual, 11/2023, A5E44914141-AC 27