Procedure
1. Use a2.5mm Allen key to loosen the retaining screw for rotation of the enclosure.
On a device with a stainless steel enclosure, loosen the front and rear retaining screw.
2. Turn the measuring cell counterclockwise to remove it from the enclosure.
Turn the measuring cell manually.
For relative pressure devices, for example, use a suitable open-ended wrench.
Ensure that the ribbon cables do not kink, pinch, strain or twist.
Take care not to damage the sensor electronics.
3. If present, remove the ribbon cable insulation .
Replacing spare parts
5.2Replacing the measuring cell
SITRANS P320/P420
Service Manual, 11/2023, A5E44914141-AC 29