Service and maintenance
6
Note
The device is maintenance-free.
6.1 Possible maintenance activities
For di󹪟cult applications (e.g. oil and gas), you can o󹪜er the following maintenance activities in
the following order:
Inspection (Page37)
Maintenance (Page38)
Analysis (Page39)
See also
List of tools (Page43)
6.1.1 Inspection
Requirement
The device is in normal operation.
You work on a closed device.
Procedure
Activity Description
Visual inspection from outside
Check the operating conditions: Ambient temperature, humidity, vibrations, dam‐
age, soiling, condensation on the display.
Check the device: Button cover, leak tightness of the process connection, electrical
connection (cable, 󹪝ttings, device plug, cable socket), grounding and equipotential
bonding.
Check whether the device is connected correctly at the process.
Check the installation conditions: Alignment of the device.
Measured value check Check whether the measured values on the display and the measured values in the
process control system match (for example: Are the values plausible and stable?)