7.4 Operating conditions
Operating conditions for gauge pressure and absolute pressure (from the gauge pressure series)
Ambient conditions
Ambient temperature
Note Observe the temperature class in hazardous areas.
Enclosure
Measuring cell with sili‐
cone oil filling
-40 … +85 °C (-40 … +185 °F)
Measuring cell with inert
oil for gauge pressure
measuring cells 1, 4, 16
and 63 bar
-40 ... +85 °C (-40 ...+185 °F)
Measuring cell with inert
oil
-20 … +85 °C (-4 … +185 °F)
Measuring cell with FDA-
compliant oil
-10 ...+85 °C (14 ... +185 °F)
Display -20 … +80 °C (-4 … +185 °F)
Storage temperature
-50 … +85 °C (-58 … +185 °F)
(for FDA-compliant oil: -20 ... + 85 °C
(-4 ... +185 °F))
Climate class in accordance
with IEC 60721-3-4
4K4H
Degree of protection accord‐
ing to IEC/EN 60529/UL50-E
Enclosure with appropriate cable
gland
IP66/Type 4X
IP68 (2 hours at 1.5 m)
Enclosure with mounted M12 con‐
nector and/or lightning protection
IP66/Type 4X
Enclosure with mounted HAN con‐
nector
IP65
Electromagnetic compatibili‐
ty
Interference emission and
interference immunity
In accordance with EN 61326 and
NAMUR NE 21
Process medium conditions
Process temperature
Technical data
7.4 Operating conditions
SITRANS P320/P420 (mA/HART)