SITRANS FM100
page 22 A5E48762204-AA
Parameter table Display
Sublevel
Parameter
level
para-
meter
para-
meter
para-
meter
Description
Value range/
value list
Default
value
LPM
Default
value
GPM
Refresh value input
Sets the display refresh
rate
0.5-10 sec 0.5 sec
Orientation
rotate CW
Rotates the display 90 °
clockwise
Landscape
rotate CCW
Rotates the display 90 °
counterclockwise
Layout
Single
Shows a measured value
in the display area
Double
Double
Shows two measured
value in the display area
display
value
Upper
display
list
selec-
tion
Sets the reading for the
upper display
Flow,
volume,
temperature,
part volume
Flow
Sets the reading for the
lower display
Temperature
Keys
Hotkeys
left
Flow
List
selection
Sets the reading and
function for the left hotkey
Off, Value,
Min / Max
Off
right
List
selection
Sets the reading and
function for the right
hotkey
Off, Value,
Min / Max
sensitivity
selec-
Sets the sensitivity for the
optical buttons
Low, Medium Low
Menu
Timeout
value input
which the setting menu is
automatically exited
without pressing a key (0
0.5 to 60 s 15 s
10.4 Measurement
The Measurement menu lists the measurement variables that the transmitter provides.
For magnetic inductive flowmeter, these are:
• Flow
• Volume (total volume counter)
• Temperature
• Part volume
Each measurement variable is still divided into its own submenu. In the submenu, all
parameters relating to the respective measuring variables can be adjusted.