SITRANS FM100
page 26 A5E48762204-AA
Parameter table Measuring
Sublevel
Para-
meter
level
Subpara-
meter level
1
Sub-
para-
meter
level 2
Sub-
para-
meter
level 3
Description
Value range/
value list
Default
value
LPM
Default
value
GPM
Flow
Unit List selection
Sets the flow unit
L/m, L/h, m3/h,
galUS/m,
galUS/h,
galUK/m,
galUK/h, L/s,
mL/s, User
L/m
galUs/
m
sepa-
ration
value input
Sets the value for the
low flow cutoff
0 ≤ Value
Range start
Range start
Simula-
tion
mode
See table Simulation mode
Volume
Counter
Type
absolute/
bidirectional
Sets the counting
mode
absolute
Unit List selection Sets the volume unit
ml, L, m3,
galUS, galUK,
User
L L
Tem-
perature
Unit List selection
Sets the temperature
unit
°C, °F, User °C °C
Simula-
tions-
modus
See table Simulation mode
Part
volume
Counter
Type
absolute/
bidirectional
Sets the counting
mode
absolute
Unit List selection Sets the volume unit
ml, L, m3,
galUS, galUK,
User
L galUs
memory
reset
Yes/No
Sets the counter
value to "0"
Simula-
tion
mode
See table Simulation mode
10.5 Dosing function
see section 12
10.6 Outputs
The FM100 flowmeter provides a total of 2 outputs that are freely configurable. The
configuration of the outputs (output 1 and output 2) is done via a wizard function.
The wizard function guides the user step by step through all necessary settings.
Steps:
Select output
Selection of the source or the measurement variable to be output (Flow,
Volume, Temperature, Part volume)
Selection of an output type (4-20 mA, 0-20 mA, 0-10 V, 2-10 V, alarm, pulse,
frequency output,IO-Link, control input)
Setting the output (scaling, thresholds)
Save the configuration