SITRANS FM100
A5E48762204-AA page 47
14. Technical Information
Measurement process: electromagnetic
Range:
see order details
Media: conductive fluids
Minimum conductivity: ≥20 µS/cm
Max. medium viscosity: 70 mm
2
/s
Max. pressure: 16 bar
Accuracy:
<±(0.8% of reading+ 0.5% of full
scale)*
Repeatability: ±0.2% of full scale
Response time flow t
90
(alarm/pulse/frequency output): <100 ms
(analogue output): <1 s
Temperature measurement
Sensor: PT1000
Accuracy: ≤±2 °C (flow >0.2 m/s)
Measuring range: temperature range of medium
Response time of temperature t
90
(signal output): <20 s
Mounting position: in all directions
In-/outlet: 3 x DN/2 x DN
Pressure drop (max at ME) MB 05 400 mbar
MB 10 50 mbar
MB 15 250 mbar (3/4 "), 30 mbar (1")
MB 20 80 mbar
MB 40 450 mbar
MB 35 120 mbar
Handling: 4 optical touch fields, useable with hand gloves
Housing: stainless steel 1.4404,
display screen PMMA
Wetted parts
Connection fitting and housing: stainless steel 1.4404
Insulation parts: PEEK
Elektrodes:
stainless steel 1.4404
Seals: FKM: MIM-12xxx; EPDM: MIM-13xxx
Protection: IP 67
Media temperature: -20 °C … +70 °C
-20 °C … +85 °C (separate version PVC cable)
-20 °C … +130 °C (separate version ETFE cable)
Ambient temperature: -20 °C … +60 °C (compact version and display
for remote version)
-20 °C…+130 °C (sensor for remote version
with ETFE cable)
-20 °C…+85 °C (sensor for remote version
with PVC cable)