page 48 A5E48762204-AA
Pulse output Push-Pull, freely scalable, configurable for
partial or accumulated totalizer
Frequency output Push-Pull, freely scalable,
Overflow frequency adjustable
Alarm output: NPN, PNP, Push-Pull,
configurable max. 30 V
DC
,
max. 200 mA short-circuit proof
Analogue output: active, 3 wire, 0(4)-20 mA,
max. load 500 Ω or 0-10 V
DC
, (R
i
= 500 Ω)
Control input (MIN / MAX / partial volume RESET)
OUT1, high active
0 <ULow <10 VDC
15 VDC <UHigh <Vs
Total and partial quantity counter: overflow (reset) at 15 * 10
6
liters
(regardless of the set unit)
Electrical connection: plug M12x1, 4-pin
Dosing Function
Dosing value: 0.1 …9999.9 [Unit]
Dosing output OUT2
PushPull, High active
Control input: OUT1, High active
0 < U
Low
< 10 V
DC
15 V
DC
< U
High
< Vs
START/STOP pulse: 0.5 s < t
high
< 4 s
RESET pulse: t
high
> 5 s
* Under reference conditions: media temperature: 15 °C ... 30 °C, 1 cSt, 500 µS/cm, 1 bar
ambience temperature: 15 °C ... 30 °C