Diagnosing and troubleshooting
9.4 Fault codes and corrective actions
SITRANS F Ultrasonic flowmeters FST030 (HART)
Operating Instructions, 08/2017, A5E35971532-AC
99
170
Loop current Loop current simulated. Disable 'Simulation'
before returning to normal operation
Only HART devices
172
Transmitter FW invalid. A component does not have the
expected FW version. Start a product firmware
update to update the component version or
replace the component.
173
Sensor FW invalid. A component does not have the
expected FW version. Start a product firmware
update to update the component version or
replace the component.
174
Display FW invalid. A component does not have the
expected FW version. Start a product firmware
update to update the component version or
replace the component.
175
IO FW invalid. A component does not have the
expected FW version. Start a product firmware
update to update the component version or
replace the component.
176
Sensor Sensor type incompatible. Please replace
sensor.
177
Device is starting Please wait until startup is finished. Startup
time see manual.
178
Transmitter FW invalid. A component does not have the
expected FW version. Start a product firmware
update to update the component version.
179 Alarm class Alarm class simulated. Disable Simulation
before returning to normal operation.
No detailed information avail-
able in view Active diagnostic
events. Icon depends on
simulated alarm class resp.
NAMUR status signal and the
priority in case of multiple