Diagnosing and troubleshooting
9.4 Fault codes and corrective actions
SITRANS F Ultrasonic flowmeters FST030 (HART)
Operating Instructions, 08/2017, A5E35971532-AC
109
ID/
Icons
Diagnostic
Action
Comment
277
LiquIdent Value above warning limit. Check process
conditions or align limit to normal operation.
Adjust parameter "Upper warning limit"
Only hydrocarbon
278
LiquIdent Value below warning limit. Check process
conditions or align limit to normal operation.
Adjust parameter "Lower warning limit"
Only hydrocarbon
279
LiquIdent Value below alarm limit. Check process condi-
tions or align limit to normal operation. Adjust
parameter "Lower alarm limit"
Only hydrocarbon
289
Sound velocity Value simulated. Disable 'Simulation' to return
to normal operation.
Only hydrocarbon and water
290
Flow velocity Value simulated. Disable 'Simulation' to return
to normal operation.
Only hydrocarbon and water
291
Pressure Value simulated. Disable 'Simulation' to return
to normal operation.
Only hydrocarbon and water
292
Kinematic viscosity Value simulated. Disable 'Simulation' to return
to normal operation.
Only hydrocarbon and water
293
Auxiliary temperature Value simulated. Disable 'Simulation' to return
to normal operation.
Only hydrocarbon and water
294
Sensor temperature Value simulated. Disable 'Simulation' to return
to normal operation.
Only hydrocarbon and water