Diagnosing and troubleshooting
9.4 Fault codes and corrective actions
SITRANS F Ultrasonic flowmeters FST030 (HART)
94 Operating Instructions, 08/2017, A5E35971532-AC
Transmitter diagnostic events
ID/
Icons
Diagnostic
Action
Comment
96
Mass flow Value above alarm limit. Check process condi-
tions or align limit to normal operation. Adjust
parameter 'Upper alarm limit'.
97
Mass flow Value above warning limit. Check process
conditions or align limit to normal operation.
Adjust parameter 'Upper warning limit'
98
Mass flow Value below warning limit. Check process
conditions or align limit to normal operation.
Adjust parameter 'Lower warning limit'
99
Mass flow Value below alarm limit. Check process condi-
tions or align limit to normal operation. Adjust
parameter 'Lower alarm limit'
100
Volume flow Value above alarm limit. Check process condi-
tions or align limit to normal operation. Adjust
parameter 'Upper alarm limit'
101
Volume flow Value above warning limit. Check process
conditions or align limit to normal operation.
Adjust parameter 'Upper warning limit'
102
Volume flow Value below warning limit. Check process
conditions or align limit to normal operation.
Adjust parameter 'Lower warning limit'
103
Volume flow Value below alarm limit. Check process condi-
tions or align limit to normal operation. Adjust
parameter 'Lower alarm limit'
104
Density Value above alarm limit. Check process condi-
tions or align limit to normal operation. Adjust
parameter 'Upper alarm limit'