WARNING
Empty pipe detection
The installation has to fulfill the following limitations for usage of the empty pipe detection
function:
● media conductivity ≥ 20 μS/cm
● length of cable at remote installation ≤ 50 (150 ft)
● special shield cable must be used
Note for MAG 1100 sizes DN 2 and DN 3:
● empty pipe detection is not available
● media conductivity ≥ 30 μS/cm
Note for MAG 5000/6000 CT (FW 3.03):
● empty pipe detection is not available
8.10 Dimensions and weight
A
1
155 (6.10)133 (5.24)
M20
(½"NPT)
D
i
L
2
A
208 (8.19)
218 (8.58)
2 x M25
B
133 (5.24)
D L
A
1
B
Table 8-14 Dimensions
Nominal size DN A
1)
A
1
/A
2
2)
B D
1
mm inch mm inch mm inch mm inch mm inch
2 1½ 161 6.34 315 12.40 186 7.33 2 0.08
3 1/8 161 6.34 315 12.40 186 7.33 3 0.12
6 ¼ 161 6.34 315 12.40 186 7.33 6 024
10 3/8 161 6.34 315 12.40 186 7.33 10 0.39
15 ½ 161 6.34 315 12.40 186 7.33 15 0.59
25 1 169 6.66 323 12.72 201 7.92 25 0.98
Technical data
8.10 Dimensions and weight
SITRANS F M MAG 1100 sensor