Loading...
Dimension-dependent batch and pulse output settings
Table A-3 50 Hz version
DN Qmax
mm Inch Min. Max.
2 1/12 - -
3 1/8 - -
6 ¼ - -
10 3/8 - -
15 ½ - -
25 1 409 μl 10.7 m
3
40 1.05 ml 27.5 m
3
50 2 1.46 ml 42.9 m
3
65 2.31 ml 72.5 m
3
80 3 3.71 ml 110 m
3
100 4 5.79 ml 172 m
3
Table A-4 60 Hz version
DN Qmax
mm Inch Min. Max.
2 1/12 0.00000095 25
3 1/8 0.00000156 41
6 ¼ 0.00000622 163
10 3/8 0.00001729 453
15 ½ 0.0000389 1019
25 1 0.000108 2833
40 0.000277 7252
50 2 0.000433 11331
65 0.000730 19149
80 3 0.00111 29007
100 4 0.00173 45325
A.2 Coil resistance
Table A-5 Coil resistance [Ω]
MAG 1100, MAG 1100F
DN Inch Resistance Tolerance
2 1/12 104 +/− 5
3 1/8 104 +/− 5
6 1/4 99 +/− 17
Appendix
A.2 Coil resistance
SITRANS F M MAG 1100 sensor
66 Operating Instructions, 01/2018, A5E03433301-AC