Installing/Mounting
4.3 Remote installation
SITRANS F M MAG 5000/6000
Operating Instructions, 01/2010, SFIDK.PS.026.G1.02
23
4. Fit and connect electrode (1) and coil (2) cables as shown in Electrical connection
(Page 34).
Note
Unscreened cable ends must be kept as short as possible.
Electrode cable and coil cable must be kept separate to prevent interference.
5. Tighten cable glands well to obtain optimum sealing.
WARNING
Mount terminal box lid before power up.