Installing/Mounting
4.3 Remote installation
SITRANS F M MAG 5000/6000
Operating Instructions, 01/2010, SFIDK.PS.026.G1.02
25
4. Mount earth wire in bottom of terminal box.
3
5
4
1
2
1 Connect electrode cable
2 Connect coil cable - keep separate from electrode cable
3 Connect power supply
4 Connect output cable
5 Connect PE (ground) wire
1. Mount connection plate in terminal box.
B
2
1
4
3
1 Electrode cable
2 Coil cable
3 Power supply
4 Output cable
1. Fit coil, electrode, supply and output cables through cable glands and connect to
connection plate as shown in Electrical connection (Page 34).
2. Fix conne
ction plate with the two diagonally opposite screws.
3. Tighten cable glands to obtain optimum sealing.
CAUTION
When remote mounted, power supply PE wire must be connected to PE terminal ( ).
Coil cable shield must be connected to SHIELD terminal.