Functions
8
This chapter describes the various menus of the transmitter in details. The menu diagrams
are shown in appendix menu diagrams (Page 73).
8.1 Output settings
Three outputs are available:
● Current output (range and time constant); terminals 31 and 32.
● Digital output (pulse, frequency, error, limit, or batch settings); terminals 56, 57, and 58.
● Relay output (error, limit, and batch settings); terminals 44, 45, and 46.
Current output
In the current output menu it is possible to select current output range and time constant, see
also Current output menu diagram (Page 79).
If c
urrent output is set to "4-20 mA+Alarm", the current output will show the following mA:
● 1.3 mA (if error level is "Fatal").
● 2 mA (if error level is "Permanent").
● 3 mA (if error level is "Warning").
For setting of error level, see error level menu diagram (Page 80).
If c
urrent output is not used, it must be set to "Off".
Digital output
Digital output can be used to configure various settings:
● Pulse (volume/pulse, pulse output, pulse width, pulse polarity, and time constant), see
pulse menu diagram (Page 79).
●
Frequency (frequency output, max frequency, and time constant), see frequency menu
diagram (Page 80).
●
Error settings (level and number), see error level menu diagram (Page 80) and error
numbe
r menu diagram (Page 80).