8.4 Reset mode
SITRANS F M MAG 5000/6000
52 Operating Instructions, 01/2010, SFIDK.PS.026.G1.02
8.4 Reset mode
The reset mode is used for resetting totalizers/counters or for restoring MAG 5000/6000 to
its factory settings.
Resetting
1. Press top up key for 2 sec.
2. Enter password.
3. Use forward key
or backward key to reach reset mode menu.
4. Press lock/unlock key
to enter reset menu.
5. Press forward key
to reach totalizer/counter to be reset or default setting menu.
6. Press lock/unlock key
to start resetting.
If restoring of factory settings is required:
1. Press lock/unlock key
again to confirm destruction of customized settings.
See also reset menu diagram (Page 85).
Zero point adjustment (MAG 6000 SV only)
Auto adjustment
Before auto zero point adjustment is carried out ensure that valves to and from flowmeter are
completely closed and that flow velocity in sensor is zero.
1. Press top up key
for 2 sec.
2. Enter password.
3. Use forward key
or backward key to reach reset mode menu.
4. Press lock/unlock key
to enter reset menu.
5. Press forward key
to reach zero adjust menu.
6. Press lock/unlock key
to enter the menu.
7. Use change key
to select "auto".
8. Press forward key
to view actual offset (lower line in display). Value will be zero after
adjustment has been performed.
9. Press lock/unlock key
to start adjustment.
Manual adjustment.
1. Press top up key
for 2 sec.
2. Enter password.
3. Use forward key
or backward key to reach reset mode menu.
4. Press lock/unlock key
to enter reset menu.
5. Press forward key
to reach zero adjust menu.
6. Press lock/unlock key
to enter the menu.
7. Use change key
to select "manual".