Technical data
11.2 Accuracy
SITRANS F M MAG 5000/6000
64 Operating Instructions, 01/2010, SFIDK.PS.026.G1.02
11.2 Accuracy
For accuracy reference conditions, please see below.
)ORZPHWHUXQFHUWDLQW\
* MAG 5100 W with DN 350 ... 2000
v ≥ 0.1 m/s (0.3 ft/s) --> E: ±0.4 ± 2 mm/s
v < 0.1 m/s (0.3 ft/s) --> E: ±(0.25/v) % of measured value
V: Flow velocity
E: Meter uncertainty as a percentage of measured value
v ≥ 0.1 m/s (0.3 ft/s) --> E: ±0.4 ± 1 mm/s *
v < 0.1 m/s (0.3 ft/s) --> E: ±(0.25/v) % of measured value
10
0,2
0,0
0,6
0,4
1,0
0,8
1,4
1,2
1,8
1,6
2,2
2,0
2,6
2,4
3,0
2,8
234 56789101112
3.30 6.69.9131620232630333639
[m/s]
[ft/s]
[±% E]
Figure 11-1 MAG 5000 with MAG 1100, MAG 1100 F, MAG 5100 W, MAG 3100 and MAG 3100 P
)ORZPHWHUXQFHUWDLQW\
V: Flow velocity
E: Meter uncertainty as a percentage of measured value
v ≥ 0.1 m/s (0.3 ft/s) --> E: ±0.2 ± 1 mm/s *
v < 0.1 m/s (0.3 ft/s) --> E: ±(0.125/v) % of measured value *
* MAG 5100 W (Order No. 7ME652...) with DN 350 ... 1200
v ≥ 0.1 m/s (0.3 ft/s) --> E: ±0.2 ± 2.5 mm/s
v < 0.1 m/s (0.3 ft/s) --> E: ±(0.25/v) % of measured value
10 234 56789101112[m/s]
[ft/s]
[±% E]
0,0
0,2
0,4
0,6
0,8
1,0
1,2
1,4
3.30 6.69.913 16 20232630 333639
Figure 11-2 MAG 6000 with MAG 1100 (not PFA), MAG 1100 F (not PFA), MAG 5100 W, MAG 3100 and MAG 3100 P