SITRANS F M MAG 5000/6000
Operating Instructions, 01/2010, SFIDK.PS.026.G1.02
91
Appendix
B
B.1 Factory settings
The transmitter is delivered with the following factory settings:
Menu item Parameter Factory
settings
Options More info
Password
Password 1000 1000 ... 9999 Changing
basic
settings
(Page 40)
Basic settings Flow direction Positive Positive, negative
Q
max
Dimension-
dependent
Dimension-dependent
•
volume unit
Dimension-
dependent
m
3
, ml, l , hl, kl, Ml, ft
3
, in
3
, US G, US kG, US MG,
US BBL, UK G, UK MG
•
time unit
Dimension-
dependent
Sec., min., hour, day
Totalizer 1 Forward Forward, reverse, net
•
Totalizer 1 unit
Dimension-
dependent
m
3
, ml, l , hl, kl, Ml, ft
3
, in
3
, US G, US kG, US MG,
US BBL, UK G, UK MG
Totalizer 2 Reverse Forward, reverse, net
•
Totalizer 2 unit
Dimension-
dependent
m
3
, ml, l , hl, kl, Ml, ft
3
, in
3
, US G, US kG, US MG,
US BBL, UK G, UK MG
Low flow cut-off 1.5% 0 ... 9.9%
Empty pipe Off On, Off
Error level Warning Fatal, permanent, warning
Basic
settings
(Page 77)
Output Current output Off On/off, Unidirectional/bidirectional, 0 ... 20 mA/4 ...
20 mA
•
Time constant
5 s
0.1 ... 30 s
Current
output
(Page 79)
Digital output Pulse Error, direction/limit, batch
1)
, frequency, pulse,
error number, off
Relay output Error Error, direction/limit, cleaning, error number, off
Direction/limit
switch
Off
1 setpoint, 2 setpoints
•
Setpoints
0 ... 100%
-100 ... +100%
•
Hysteresis
5%
0.0 ... 100%
Direction/lim
it (Page 81)
Batch
1)
Off
•
Batch quantity
0
Dimension-dependent
Batch
(Page 82)