MAG 8000 & MAG 8000 CT
20
SFIDK.PS.026.D9.02
4. Commissioning
Output configuration in
Flow Tool (PC software)
The Flow Tool window shows the parameters for output configuration. Each parameter has its
own guideline in selecting the correct parameter setting.
Pulse output, volume
selection (MAG 8000)
Max. Guidelines for min. volume per pulse at Q
n
flow
Volume [m
3
] = Q
n
[m
3
/s] * (2*PW [s])
DN rate Qn 5 ms 10 ms 50 ms 50 ms 50 ms 100 ms 500 ms
(Inches) (Q3) PW m
3
PW m
3
PW m
3
PW gallon PW Ml PW m
3
PW m
3
m
3
[50Hz] [50Hz] [10Hz] [10Hz] [10Hz] [5Hz] [1Hz]
25 (1") 17.67 0.00005 0.0001 0.0005 0.130 0.000001 0.001 0.005
40 (1½“) 45 0.0001 0.0003 0.001 0.330 0.000001 0.003 0.013
50 (2") 63 0.0002 0.0004 0.002 0.462 0.000002 0.004 0.018
65 (2½“) 100 0.0003 0.0006 0.003 0.734 0.000003 0.006 0.028
80 (3") 160 0.0004 0.0009 0.004 1.174 0.000004 0.009 0.044
100 (4") 250 0.0007 0.0014 0.007 1.835 0.000007 0.014 0.069
125 (5") 400 0.0011 0.0022 0.011 2.935 0.000011 0.022 0.111
150 (6") 630 0.0018 0.0035 0.018 4.623 0.000018 0.035 0.175
200 (8") 1000 0.0028 0.0056 0.028 7.338 0.000028 0.056 0.278
250 (10") 1600 0.0044 0.0089 0.044 11.741 0.000044 0.089 0.444
300 (12") 2500 0.0069 0.0139 0.069 18.345 0.000069 0.139 0.694
350 (14") 3463 0.0096 0.0192 0.096 25.412 0.000096 0.192 0.962
400 (16") 4523 0.0126 0.0251 0.126 33.190 0.000126 0.251 1.256
450 (18") 5725 0.0159 0.0318 0.159 42.010 0.000159 0.318 1.590
500 (22") 7068 0.0196 0.0393 0.196 51.865 0.000196 0.393 1.963
600 (24") 10178 0.0283 0.0565 0.283 74.687 0.000283 0.565 2.827
PW = pulse width
Note
Display volume for 5 ms pulse width is based on a basic version with maximum 50 Hz pulse output
rate.
For the advanced version, with maximum 100 Hz pulse rate, the pulse volume values can be
reduced to half.
The calculated numbers of pulses are an average of the measuring period.
For MAG 8000 CT please see definition of Q
3
flow rates in the certificates for OIML R49 or MI 001.