Net flow output
The MAG 8000 has a special net pulse output that includes bi-directional flow calculations.
The example shows that over time, the net pulse output indicates the bi-directional totalizer as
calculated internally.
The same principle applies for forward and reverse flow calculation. By changing the status of
the pulse output, the internal pulse calculator will be reset.
Net totalizer in
meter display
(Bi-directional)
Pulse output forward
Uni-directional mode
Volume [m
3
]
Pulse output net forward
Bi-directional mode
Volume [m
3
]
Flow
Volume [m
3
]
Delivered
volume
Internal
calculation
Total accounted volume
[m
3
]
Forward/Reverse
Delivered
volume
Internal
calculation
0-000
10 - 10 0 10
−2- 0−12 0
18 - 20 −12+20= 8
18F 30F 18F
FT ID Meter Display Parameter/data type Factory settings Data range
number version view Fix parameter or meter data that not are changeable
1 All M1 Application identifier Indetity Max. 14 characters. Only numbers
are visible on the display
2 All - Application location Location Max. 14 characters
3 All M3 Module type MLFB depended Basic or Advanced
4 All M4 Software version x.xxPxx (x.xxPx.x)
5 All - Sensor size Sensor related DN 25..600(1"..24")
6 All - Vendor name Siemens Siemens
8 All - Totalizer unit MLFB depended Max. 10 characters
9 All - Flowrate unit MLFB depended Max. 10 characters
10 All - Qn (Q3) Sensor related 0 to 1*10^9
11 All - Product code number 7ME6810XXXXXXXXX
7ME681 or 7ME682 XXXXXXX
12 All - Serial number XXXXXXNXXX
100 All M2 Actual date and time PS3 production date and time year-month-day T
hours:minutes:seconds
101 All O1 Totalizer 1 0 0 to +-2*10^9
102 All O2 Totalizer 2 0 0 to +-2*10^9
103 All O5 Customer totalizer 3 0 0 to +-2*10^9
104 All O5 Reset customer totalizer 3 No Yes/no
4.3 Parameter list MAG 8000 is delivered with factory settings that are not stored as default values. Because defaults
values are not present in the meter, an automatic return to factory values is not possible.
The default settings are available at
www.siemens.com/flow. Navigate to Tools & Downloads
under MAG 8000.
Visible display information is indicated in the table by menu and index number. Remember to
enable displayed menus FT130.
The abbreviations used in the display menu table are: Operator menu = O, Meter menu = M,
Service menu = Se, Data Logger menu = L, Statistic menu = St, Revenue menu = R.