MAG 8000 & MAG 8000 CT
44
SFIDK.PS.026.D9.02
Transmitter
Installation MAG 8000 Integral (compact) or remote with factory mounted cable in
5, 10, 20 or 30 m lengths with IP68/NEMA 6P connectors.
Connection is made at the transmitter bottom.
Installation MAG 8000 CT Integral (compact) or remote with factory mounted cable in 5 or
10 m lengths with IP68/NEMA 6P connectors.
Connection is made at the transmitter bottom.
MI-001 verified meters only Integral (compact) mounted.
Enclosure Stainless steel top housing (AISI 316) and coated brass bottom.
Remote wall mount bracket in stainless steel (AISI 304).
Cable entries 2 x M20 (one gland for one cable of size 6 … 8 mm
(0.02 ... 0.026 ft) is included in the standard delivery)
Display and key Display with 8 digits for main information.
Index, menu and status symbols for dedicated information
Key for toggling through the information and reset customer
totalizer and call-up function
Selectable default information and accessible menus:
- Operator
- Meter
- Service
- Data Logger
- Statistic and leakage (only Advanced version)
- Revenue and Tariffs (only Advanced version)
Totalized information can be displayed with 1, 2, 3 decimals or
automatic adjustment for maximum resolution
Flow unit MAG 8000
Europe std. Volume in m
3
and flow rate in m
3
/h
US std. Volume in Gallon and flow rate in GPM
Australian std. Volume in Ml and flow rate as Ml/d
Other units selectable:
Volume: m
3
x 100, l x 100, G x 100, G x 1000, MG,
CF x 100, CF x 1000, AF, Al, kl
Flow: m
3
/min, m
3
/d, l/s, l/min, GPS, GPH, GPD, MGD, CFS,
CFM, CFH
Other units are ordered from factory or manually configured onsite
by sticking a label on the display and changing the scaling factors
Flow unit MAG 8000 CT
Europe std. Volume in m
3
and flow rate in m
3
/h
Digital output MAG 8000 2 passive outputs (MOS), individual galvanically isolated
Maximum load ± 35 V DC, 50 mA short circuit protected
Output A function
Programmable as pulse volume
– forward – reverse – forward/net
– reverse/net
Output B function
Programmable as pulse volume (like output A), alarm or call-up
Output
Max. pulse rate of 50 Hz (only Basic version) and 100 Hz (only
Advanced version), pulse width of 5, 10, 50, 100, 500 ms
Digital output MAG 8000 CT 2 passive outputs (MOS), individual galvanically isolated
Maximum load ± 35 V DC, 50 mA short circuit protected
Output A function
Programmable as pulse volume
– forward
Output B function
– Alarm
Output
Max. pulse rate of 50 Hz (only Basic version) and 100 Hz (only
Advanced version), pulse width of 5, 10, 50, 100, 500 ms
8. Technical data
(continued)