MAG 8000 & MAG 8000 CT
16
SFIDK.PS.026.D9.02
4. Commissioning
Select the parameter or group which is to be read, written, printed, or exported to a CSV file by
selecting via the right mouse button in left or right window.
Only data with white background can be changed. Red text indicates off-line data. Black text
shows data identical with meter data.
Each parameter has a prompt advising the purpose of the parameter and the setting limitations.
Select a parameter, by clicking in the white cell in the right window of the Flow Tool.
Depending on the parameter selected, a form or dialog box will open to allow selections or data
entry. The figure shows the alarm status, where marked alarms are enabled.
Read - write, print or
export the meter data
Data protection by
hardware key
A hardware key is installed in the HL hole to
change protected parameters. The HL hole is
located in the front of the PCB board behind the
battery. (FT = Flow Tool parameter number)
Protected parameters are:
New Password
FT5 - Sensor tube diameter
FT7 - Meter No.
FT8 - Totalizer unit
FT9 - Flow unit
FT10 - Q
n
(Q
3
)
FT300 - Totalizer unit factor
FT301 - Flow unit factor
FT302 - Pipe size
FT321 - Calibration date
FT323 - Calibration factor
FT325 - Sensor offset
FT332 - Max. sensor excitation frequency
Additional protected CT parameters:
FT101 - Totalizer 1
FT102 - Totalizer 2
FT303 - Operation excitation frequency
FT305 - Decimal point place
FT310 - Flow direction totalizer 1
FT312 - Flow direction totalizer 2
FT327 - Adjustment factor
FT328 - Low flow cut off
FT332 - Empty pipe impedance
FT550 - Coil current active
FT551 - Fix flow mode active