SITRANS F M MAGFLO
®®
®®
®
Electromagnetic flowmeter type MAG 5100 W
Installation, general
Reading and operating the flowmeter is possi-
ble under almost any installation conditions
because the display can be oriented in relation
to the sensor. To ensure optimum flow mea-
surement attention should be paid to the follow-
ing:
The sensor must always be completely full with
liquid.
Therefore avoid:
Installation at the highest point in the pipe
system
Installation in vertical pipes with free outlet
For partially filled pipes or pipes with downward
flow and free outlet the flowmeter should be
located in a U-tube.
Installation in vertical pipes
Recommended flow direction: upwards. This
minimizes the effect on the measurement of
any gas/air bubbles in the liquid.