Param‐
eter ID
Meter
version
Display
view
Parameter/data type Factory settings Data range
Fixed parameter or meter data that are not changeable
6 All - Vendor name Siemens Siemens
8 All - Totalizer unit MLFB dependent Max. 12 characters
9 All - Flowrate unit MLFB dependent Max. 12 characters
10 All - Qn (Q3) Sensor related 0 to 1x10
9
11 All - Product code number 7ME6810, 7ME6820 or
7ME6880
12 All - Serial number XXXXXXHXXX Max. 12 characters
C.2.2 100-199
Param‐
eter ID
Meter
version
Display
view
Parameter/data type Factory settings Data range
Fixed parameter or meter data that are not changeable
100 All M2 Actual date and time PS3 production date and
time
year-month-day T
hours:minutes:seconds
101 All O1 Totalizer 1 0 0 to ±2x10
9
102 All O2 Totalizer 2 0 0 to ±2x10
9
103 All O5 Customer totalizer 3 0 0 to +±2x10
9
104 All O5 Reset customer totalizer 3 No Yes/No
This parameter is removed from
PDM parameter list. The func‐
tionality is implemented via
PDM menu "Device" → "Totaliz‐
er" → "Reset customer Totalizer
3".
105 All - Customer totalizer 3 reset
date
PS3 production date and
time
year-month-day T hours:mi‐
nutes:seconds
106 All - Flow rate 0 to 1.25 Qn (Q3)
107 All - Actual velocity Measured value ±15 m/s
108 All - Flowrate percent value 0 to 125% (Q4)
Appendix
C.2 Parameter lists
SITRANS MAG 8000