Param‐
eter ID
Meter
version
Display
view
Parameter/data type Factory settings Data range
Fixed parameter or meter data that are not changeable
274 All - High flow alarm fault disap‐
pears
2000-01-01 T 00:00:00
275 All - Reverse high flow alarm out‐
put enable
No Yes / No
276 All - Reverse high flow alarm fault
hours
0
277 All - Reverse high flow alarm fault
counter
0
278 All - Reverse high flow appears 2000-01-01 T 00:00:00
279 All - Reverse high flow disappears 2000-01-01 T 00:00:00
C.2.4 300-399
Param‐
eter ID
Meter
version
Display
view
Parameter/data type Factory settings Data range
Fixed parameter or meter data that are not changeable
300 All - Totalizer volume unit factor MLFB-dependent 0 to 1*x0
10
301 All - Flow unit factor MLFB-dependent 0 to 1*x0
10
302 All - Pipe size Sensor-related 25 to 1200
303 All - Meter excitation frequency
(in battery power mode)
1/15 Hz 1/60 Hz, 1/30 Hz, 1/15 Hz, 1/5 Hz,
1.5625 Hz, 3.125 Hz, 6.25 Hz
304 All - Mains frequency MLFB-dependent 50 or 60 Hz mains
305 All - Decimal point MLFB-dependent No point, One digit after point, Two
digits after point, Three digits after
point, Automatic point adjust
306 All - Displayed unit Use m
3
unit Displayed m
3
unit or an arrow indica‐
tion for a unit label
307 All - Measurement direction MLFB-dependent Bidirectional, Forward only, Reverse
only
310 All - Flow direction totalizer 1 Forward Forward, Reverse or Bi-directional
net flow
311 All - Totalizer 1 changes date PS3 production date
and time
312 All - Flow direction totalizer 2 MLFB-dependent Forward, Reverse or Bi-directional
net flow
313 All - Totalizer 2 changes date PS3 production date
and time
320 All - Verification mode enable No Yes / No
Value is reset to "No" when upload‐
ing device parameters to pc after four
hours
321 All - Calibration date Calibration date year-month-day T hours:mi‐
nutes:seconds
Appendix
C.2 Parameter lists
SITRANS MAG 8000
142 Operating Instructions, 11/2018, A5E03071515-AE