Param‐
eter ID
Meter
version
Display
view
Parameter/data type Factory settings Data range
Fixed parameter or meter data that are not changeable
323 All - Calibration factor Sensor-related
324 All - Gain correction Sensor-related
325 All - Sensor offset Sensor-related
327 All - Adjustment Factor 1 -2 to 2
328 All - Low flow cut-off MAG 8000 Standard
(7ME6810) 0.05 % for
FW 3.07 and lower. For
FW 3.09 see table C1
MAG 8000 Standard pa‐
rameter 328 FW 3.09 be‐
low.
MAG 8000 CT
(7ME6820) 0.25 % for
FW 3.07 and lower. For
FW 3.09 see table C2
MAG 8000 CT parame‐
ter 328 FW 3.09 below.
MAG 8000 Irrigation
(7ME6880) 1 %
0 to 9.9%
329 All - Filter time constant 5 Tau 1 to 1000
331 All - Excitation frequency limit 6.25 Hz for advanced
version and 1/15 Hz for
basic version
332 All - Excitation frequency sensor
limit
Sensor-related 7ME6810 and 7ME6820:
6.25 Hz
(DN 25 to 200 (1" to 8"))
3.125 Hz
(DN 250 to 600 (10" to 24"))
1.5625 Hz
(DN 700 to 1200 (28" to 48"))
7ME6880:
3.125 Hz
(DN 25 to 600 (1" to 24"))
1.5625 Hz
(DN 700 to 1200 (28" to 48"))
333 All - Empty pipe detection ena‐
ble
Yes Yes / No
334 All - Empty pipe limit 25000 ohm 0 to 2.15x10
9
Table C-1 MAG 8000 Standard (7ME6810) parameter 328 FW 3.09
Sensor size Parameter value
DN 25 (1") 0.16
DN 40 (1½") 0.15
DN 50 (2") 0.17
Appendix
C.2 Parameter lists
SITRANS MAG 8000
Operating Instructions, 11/2018, A5E03071515-AE 143