2-45
1010PFM-3G
Section 2
Press and the highlight returns to the [Point #] field.
To enable numeric entry press then type [2] (for the second calibration point).
Press and the cursor moves to the point # 2 [Indicated Rate] field.
Use the numeric keys to type a flow rate for point # 2. The value entered here must
be of a greater magnitude than the value entered for point #1.
Press and the cursor moves to the [% Correction] field.
To enable numeric entry press
Use the numeric keys to type the required % correction for point # 2 then press
Repeat this procedure for all required calibration points. After entering the last Flow
Calibration Point press to exit MultiPoint calibration.
ENT
ENT
ENT
ENT
2.8 THE STRIPCHART SETUP MENU
This menu becomes available after picking a Meter Type and Measurement Channel. Use Stripchart
Setup to select a data category, the type of scaling and the update interval for the on-screen stripchart
display. You can also clear the stripchart screen.
Stripchart Setup Select Data Vfo
Vfab
Temp
Vs
Valc
Vaer
Data Display Data Rate Units
Percent of Span
Time Base 1 Second
3 seconds
6 seconds
12 seconds
24 seconds
1 minute
3 minutes
6 minutes
12 minutes
24 minutes
1.2 hours
Stripchart Clear Yes
No
THE STRIPCHART SETUP MENU STRUCTURE