2-53
1010PFM-3G
Section 2
2.10.1 ANALOG OUT SETUP
System 1010 provides current, voltage and pulse-rate analog outputs. Analog Out Setup allows you to
assign data functions for these signals. The flow computer’s terminal strip contains the analog output
terminals.
Vfo Spanned volumetric/mass flow rate
Vfab Spanned unsigned flow magnitude
Vs Spanned liquid sonic velocity
Valc Receive signal amplitude
Vaer Relative degree of liquid aeration/cavitation
T1 Current liquid temperature
Analog Out Setup Data Categories
I/O Data Control Analog Output Setup Io1 Vfo
Vfab
Vs
Valc
Vaer
T1
Vo1 *See Io1 option list
Pgen1 *See Io1 option list
Relay Setup Relay 1 Off
Power On
Hi Temperature
Low Temperature
High Flow
Low Flow
Flow Alarm
Fault Alarm
Soft Fault
Spacing
Empty
Aeration
Interface
Reverse Flow
Batch Tot
Pos Total
Neg Total
Relay 2 *See Relay 1 list
Analog Input Setup Iin1 Input Off
Aux
4 mA *numeric entry
20 mA *numeric entry
Vin1 Input Off
Aux
0 Volts *numeric entry
10 Volts *numeric entry
THE I/O DATA CONTROL MENU STRUCTURE
Ú
Ú
Ú
Ú
Ú
Ú
Ö
Ö
Ö
Ö
Ö
Ö
Ö
Ö
Ö
Ö
Ö
Ö
Ö
Ö
Ö
Ö
ÖÖ
Ö
Ö
Ö
Ö
Ú
Ú
Ú
Ú
(Note: For Dual Path flowmeters: “1” = Path 1, “2” = Path 2 and “S” represents the system or average channel. These
characters appear to the left of the option list parameter.)
Io (Isolated current) 4 to 20 mA varies in proportion to an assigned data function.
Vo (DC voltage) 0 to 10 Vdc varies in proportion to an assigned data function.
Pgen (TTL logic) 0 to 5000 Hz varies in proportion to an assigned data function.
System 1010 Analog Outputs