2-61
1010PFM-3G
Section 2
2.11.3 THE APPLICATION INFO MENU
This menu provides a live display of the basic timing data used by meter during operation.
2.11.4 THE LIQUID DATA MENU
This menu shows the current Reynolds number used by the meter to implement flow profile compen-
sation.
NOTE: On some 1010 models, the Liquid Data Menu screen also displays current RTD temp-
erature(s).
TN uSec 75.193
TL uSec 44.009
DeltaT uSec 126
Burst/Sec 256
% Accept 100
Last Makeup (Re)Start
Makeup Status Measurement
Siemens 2 Channel [1] 1
Time between transit and receive
Application Info
TN uSec The total elapsed time between the transmission and reception of a
sonic pulse.
TL uSec The total time a sonic pulse takes to travel through the liquid.
DeltaT nSec This is the instantaneous result of subtracting the down from the up tran-
sit times.
Note: This will appear to be more active than dampened flow
readings.
Burst/Sec Number of transmissions per second under current operating conditions.
The following factors influence this parameter: aeration, pressure trans-
ients, signal strength variation.
% Accept The % of accepted up/down transmit burst sets.
Last Makeup Reason for last Makeup (signal reacquisition).
Makeup Status Current stage of Makeup routine. [Measurement = normal operation]
APPLICATION INFO MENU ITEMS
Reynolds # 318017
Siemens 2 Channel [1] Channel 1
Current Reynolds number
Liquid Data